Reviewing a reported concentration: records to request
A five-part checklist for following a reported value back through its dataset, analytical run and calculation records.
You have a concentration in a report table and need to understand how it was produced. Asking for “all the raw data” can leave both the laboratory and the reviewer unclear about what is needed. Start by naming the result and the question you want to resolve.
Define the result first
Record the study and report version, the sample identifier, the analyte, the reported value and its unit. Work within authorised access arrangements; do not send participant or client records through a public website.
Request the records that connect it
- The reported dataset. Identify the export revision and the row used in the report or supplied for pharmacokinetic analysis. Check sample identifiers, units and any flags.
- The sample’s analytical history. Request the run reference and relevant preparation or dilution records. Establish whether the reported value came from the first analysis or a later analysis.
- The run’s supporting records. Locate calibration and quality-control results, the documented run assessment and any investigation relevant to that sample.
- The processing and calculation record. Identify the method version, processing settings and factors needed to explain the value. A spreadsheet total alone may not show those decisions.
- Changes and decisions. Where applicable, request the reason for reanalysis or reprocessing, the values retained, and the decision supporting the reported result.
Leave a review someone else can follow
Record the evidence references, what you compared, any difference and what remains unavailable. A difference is a question to investigate; it does not, by itself, establish an error or invalidate a study.
Sources